RecallRadar

FDA Device recall Z-0070-2026

Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version:…

On 2025-10-29 the FDA classified a Class I recall of Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version:… by B Braun Medical, citing the potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0070-2026
ClassificationClass I
Statusongoing
Initiated2025-09-30
Published2025-10-29
Last updated2026-09-13 11:57 UTC
CompanyB Braun Medical
DistributionUS

Product

Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version: N/A Product Description: STREAMLINE BLOODLINE SET FOR DIALOG Component: N/A

Reason

The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.

Identifiers

code_info
Model No. SL-2010M2096; UDI-DI 04046964367786; Lot Numbers: A2500119, A2500121, A2500128, A2500129, A2500130, A250013…Model No. SL-2010M2096; UDI-DI 04046964367786; Lot Numbers: A2500119, A2500121, A2500128, A2500129, A2500130, A2500132, A2500133, A2500134, A2500149, A2500150, A2500151, A2500180, A2500181, A2500182, A2500185, A2500198, A2500199, A2500200, A2500203, A2500204, A2500205, A2500206, A2500207, A2500208, A2500209, A2500210, A2500211, A2500212, A2500213, A2500214, A2500215, A2500216, A2500217, A2500218, A2500219, A2500220, A2500246, A2500247, A2500248, A2500249, A2500250, A2500251, A2500257, A2500258, A2500259, A2500260, A2500261, A2500262, A2500267, A2500268, A2500269, A2500273, A2500274, A2500275, A2500276, A2500277, A2500278, A2500295, A2500298, A2500299, A2500300; ***Addded 3/24/2026***Additional Lots Identified A2500318, A2500319, A2500320, A2500321, A2500322, A2500323, A2500324, A2500325, A2500326, A2500327, A2500328, A2500329, A2500330, A2500356, A2500357;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0070-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.