On 2021-10-13 the FDA classified a Class II recall of CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with heat exchange devices by Cardioquip, citing there is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potenti….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0069-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-06-24 |
| Published | 2021-10-13 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Cardioquip |
| Distribution | US |
Product
CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with heat exchange devices.
Reason
There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potentially aerosolized into the operating room during surgery which could result in patient infection.
Identifiers
| code_info | All Serial Numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0069-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.