RecallRadar

FDA Device recall Z-0066-2025

DiamondTOUCH Digital Inflation Device, REF: IN9135/A, IN9152/A

On 2024-10-23 the FDA classified a Class II recall of DiamondTOUCH Digital Inflation Device, REF: IN9135/A, IN9152/A by Merit Medical Systems, citing inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which could lead to infection.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0066-2025
ClassificationClass II
Statusongoing
Initiated2024-08-19
Published2024-10-23
Last updated2026-09-13 11:57 UTC
CompanyMerit Medical Systems
DistributionUS

Product

DiamondTOUCH Digital Inflation Device, REF: IN9135/A, IN9152/A

Reason

Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which could lead to infection.

Identifiers

code_infoREF/UDI-DI/Lot(Expiration): IN9135/A/00884450409640/H2918441(4/15/2027). IN9152/A/00884450409657/H2912682(4/8/2027)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0066-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.