FDA Device recall Z-0064-2019

Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory…

On 2018-10-10 the FDA classified a Class II recall of Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory… by Leica Microsystems, citing these devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0064-2019
ClassificationClass II
Statusterminated
Initiated2018-06-13
Published2018-10-10
Last updated2026-09-13 11:54 UTC
CompanyLeica Microsystems
DistributionUS

Product

Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.

Reason

These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.

Identifiers

code_infoSerial Numbers: 70-78, 111-112, 120-175

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.