FDA Device recall Z-0059-2014

Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM1.S Rx Only, Manufacturer GMS- Gesellschaft fur medizinisch…

On 2013-10-30 the FDA classified a Class II recall of Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM1.S Rx Only, Manufacturer GMS- Gesellschaft fur medizinisch… by Integra Lifesciences, citing integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0059-2014
ClassificationClass II
Statusterminated
Initiated2013-06-11
Published2013-10-30
Last updated2026-09-13 11:53 UTC
CompanyIntegra Lifesciences
DistributionUS

Product

Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM1.S Rx Only, Manufacturer GMS- Gesellschaft fur medizinische, Sondentechnik mbH Dofstrasse 2 24247 Melkendorf, Germany Distributed by Integra Neurosciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Used in direct monitoring of the partial pressure of oxygen.

Reason

Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.

Identifiers

code_infoK002765 Lot # 090812, 170812

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.