RecallRadar

FDA Device recall Z-0058-2018

THERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-S

On 2017-11-08 the FDA classified a Class II recall of THERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-S by Biosense Webster, citing biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand cathe….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0058-2018
ClassificationClass II
Statusterminated
Initiated2017-09-06
Published2017-11-08
Last updated2026-09-13 11:54 UTC
CompanyBiosense Webster
Distributionn/a

Product

THERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-S

Reason

Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.

Identifiers

code_info
17637495L 17638389L 17639706L 17639707L 17639708L 17640039L 17640494L 17640952L 17640040L 17640742L 17640…17637495L 17638389L 17639706L 17639707L 17639708L 17640039L 17640494L 17640952L 17640040L 17640742L 17640953L 17640041L 17640743L 17640954L 17640490L 17640744L 17643130L 17638118L 17638387L 17638388L

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0058-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.