RecallRadar

FDA Device recall Z-0053-2025

Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket

On 2024-10-16 the FDA classified a Class II recall of Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket by Medline Industries Northfield, citing CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0053-2025
ClassificationClass II
Statusongoing
Initiated2024-08-28
Published2024-10-16
Last updated2026-09-13 11:57 UTC
CompanyMedline Industries Northfield
DistributionUS

Product

Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket

Reason

CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues, organs, and calculi during general and laparoscopic procedures. The tube inside the pouch, which detaches during the removal process, may stretch out from the pouch opening if it is not precisely fixed during the production process. Medline has identified that this product is utilized as a component within the Medline-branded convenience kits.

Identifiers

code_infoLot numbers 24AMA209, exp. 2025-06-30; 24BMB025, exp. 2025-08-31; 23LMA745, exp. 2025-07-31; UDI-DI each 10195327384494, UDI-DI case-40195327384495

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0053-2025%22