FDA Device recall Z-0048-2014
STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is manufactured and distributed b…
- FDA Device
- Class II
- terminated
- Published 2013-10-30
On 2013-10-30 the FDA classified a Class II recall of STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is manufactured and distributed b… by Abbott Medical Optics, citing this corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0048-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-09-16 |
| Published | 2013-10-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Abbott Medical Optics |
| Distribution | US |
Product
STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.
Reason
This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).
Identifiers
| code_info | All units of these models: STAR S2 Excimer Laser System; STAR S3 ActiveTrak; STAR S4 with Variable Spot Scanning; and S…All units of these models: STAR S2 Excimer Laser System; STAR S3 ActiveTrak; STAR S4 with Variable Spot Scanning; and STAR S41R; There are no expiration dates for the STAR Excimer Laser Systems. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0048-2014%22