Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer

On 2019-10-16 the FDA classified a Class II recall of Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer by Ohmeda Medical, citing GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris ar….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0041-2020
ClassificationClass II
Statusterminated
Initiated2018-10-05
Published2019-10-16
Last updated2026-09-13 11:55 UTC
CompanyOhmeda Medical
DistributionUS

Product

Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Reason

GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.

Identifiers

code_infoGiraffe OmniBeds shipped from 2000 to 2011 with Serial Numbers starting from HDGD00001 to HDGQ50183.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0041-2020%22