RecallRadar

FDA Device recall Z-0040-2024

Citrine Triple Quad Mass Spectrometer, Part Number 5063684

On 2023-10-11 the FDA classified a Class II recall of Citrine Triple Quad Mass Spectrometer, Part Number 5063684 by Ab Sciex, citing in rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0040-2024
ClassificationClass II
Statusongoing
Initiated2023-08-30
Published2023-10-11
Last updated2026-09-13 11:56 UTC
CompanyAb Sciex
DistributionUS

Product

Citrine Triple Quad Mass Spectrometer, Part Number 5063684

Reason

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Identifiers

code_info
UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901,…UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0040-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.