Citrine QTRAP Mass Spectrometer, Part Number 5063685

On 2023-10-11 the FDA classified a Class II recall of Citrine QTRAP Mass Spectrometer, Part Number 5063685 by Ab Sciex, citing in rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0039-2024
ClassificationClass II
Statusongoing
Initiated2023-08-30
Published2023-10-11
Last updated2026-09-13 11:56 UTC
CompanyAb Sciex
DistributionUS

Product

Citrine QTRAP Mass Spectrometer, Part Number 5063685

Reason

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Identifiers

code_info
UDI-DI: 00842365100540; Serial Numbers: DD250051901, DD250172009, DD250182009, DD230001804, DD250111910, DD250132004, D…UDI-DI: 00842365100540; Serial Numbers: DD250051901, DD250172009, DD250182009, DD230001804, DD250111910, DD250132004, DD250091909, DD250101910, DD250192010, CN10011707PT