FDA Device recall Z-0038-2020

V-Twin, Catalog # 6002-800, UDI: 03661540600180 - Product Usage: An in vitro diagnostic device, which is an automated chemistry analyzer, m…

On 2019-10-16 the FDA classified a Class II recall of V-Twin, Catalog # 6002-800, UDI: 03661540600180 - Product Usage: An in vitro diagnostic device, which is an automated chemistry analyzer, m… by Vital Scientific N V, citing the action is being initiated because the cooling indicator board was improperly installed. This was first noticed by a service employee during assembly. If for some reason the co….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0038-2020
ClassificationClass II
Statusterminated
Initiated2018-09-07
Published2019-10-16
Last updated2026-09-13 11:55 UTC
CompanyVital Scientific N V
DistributionUS

Product

V-Twin, Catalog # 6002-800, UDI: 03661540600180 - Product Usage: An in vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.

Reason

The action is being initiated because the cooling indicator board was improperly installed. This was first noticed by a service employee during assembly. If for some reason the cooling of the reagent rotor fails, the reagent cooling indicator light will not illuminate and the acoustic beeper will not go off. As a consequence the operator will not be warned and the reagent storage conditions may fall out of specification.

Identifiers

code_infoThe following serial numbers: 14-3301-14-3344; 15-3301-15-3326; 16-3301 -16-3353; 17-3301 -17-3346; 18-3301-18-3323

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.