RecallRadar

FDA Device recall Z-0036-2026

GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040…

On 2025-10-08 the FDA classified a Class II recall of GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040… by W L Gore And Associates, citing reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so I….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0036-2026
ClassificationClass II
Statusongoing
Initiated2025-08-26
Published2025-10-08
Last updated2026-09-13 11:57 UTC
CompanyW L Gore And Associates
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY

Product

GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A, ECH060040W, ECH060050A, ECH060050J, ECH460045A, ECH460045J, ECH470045, ECH470045A; used with Instructions for Use.

Reason

Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.

Identifiers

code_info
REF/UDI-DI: ECH050020J/04993024010758, ECH050020W/00733132635450, ECH050050J/04993024010772, ECH050050W/00733132635474…REF/UDI-DI: ECH050020J/04993024010758, ECH050020W/00733132635450, ECH050050J/04993024010772, ECH050050W/00733132635474, ECH060010A/00733132622573, ECH060020A/00733132622597, ECH060020J/04993024009967, ECH060020W/00733132634729, ECH060040/00733132605347, ECH060040A/00733132605354, ECH060040W/00733132619719, ECH060050A/00733132626960, ECH060050J/04993024009981, ECH460045A/00733132634811, ECH460045J/04993024010789, ECH470045/00733132626977, ECH470045A/00733132634811. Product with a serial number that includes and falls within the range of 6597081PP001 and 9853822PP024. Instructions for Use MD183888 Revision 4

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0036-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.