FDA Device recall Z-0034-2013
CYTO-STAT tetraCHROME CD45- FITC/CD56-RD1/CD19-ECD/CD3-PC5, Part Number 6607073
- FDA Device
- Class II
- terminated
- Published 2012-10-17
On 2012-10-17 the FDA classified a Class II recall of CYTO-STAT tetraCHROME CD45- FITC/CD56-RD1/CD19-ECD/CD3-PC5, Part Number 6607073 by Beckman Coulter, citing the recall was initiated because Beckman Coulter has determined that the specimen and prepared sample stability claims stated in the product labeling for tetraCXP software and CYT….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0034-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-05-22 |
| Published | 2012-10-17 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Beckman Coulter |
| Distribution | US |
Product
CYTO-STAT tetraCHROME CD45- FITC/CD56-RD1/CD19-ECD/CD3-PC5, Part Number 6607073. Allows simultaneous identification and enumeration of total CD3+, total CD4+, total CD8+, dual CD3+/CD4+, dual CD3+/CD8+ and/or total CD3+, CD19+ and CD3+/CD56+ lymphocyte percentages and absolute counts in whole blood by flow cytometry. The systems also provide the CD4/CD8 ratio when using CD45-FITC/ CD4-RD1/CD8-ECD/CD3-PC5, and total lymphocyte percentage when using CD45-FITC/CD56-RD1/ CD19-ECD/CD3-PC5.
Reason
The recall was initiated because Beckman Coulter has determined that the specimen and prepared sample stability claims stated in the product labeling for tetraCXP software and CYTO-STAT reagents (CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5) need to be updated for alignment with current standards and clinical decision-making associated with lymphocyte immunophenotyping
Identifiers
| code_info | All tetraCHROME lots manufactured |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0034-2013%22