FDA Device recall Z-0033-2022

Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) - Product Usage: use in the calibration of the Alinity c CRP Vari…

On 2021-10-13 the FDA classified a Class III recall of Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) - Product Usage: use in the calibration of the Alinity c CRP Vari… by Sentinel Ch Spa, citing calibrator vials were incorrectly labeled. Use of incorrectly labeled vials will cause the instrument to generate a warning that calibrators are incorrectly loaded and calibration….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0033-2022
ClassificationClass III
Statusterminated
Initiated2021-07-29
Published2021-10-13
Last updated2026-09-13 11:56 UTC
CompanySentinel Ch Spa
DistributionUS

Product

Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) - Product Usage: use in the calibration of the Alinity c CRP Vario assay, Cardiac CRP high sensitivity method [Cardiac CRP], on the Alinity c analyzer.

Reason

Calibrator vials were incorrectly labeled. Use of incorrectly labeled vials will cause the instrument to generate a warning that calibrators are incorrectly loaded and calibration will not be completed successfully.

Identifiers

code_infoLot 00681Y600

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.