FDA Device recall Z-0032-2026

DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352

On 2025-10-08 the FDA classified a Class II recall of DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352 by Beckman Coulter, citing on a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0032-2026
ClassificationClass II
Statusongoing
Initiated2025-08-04
Published2025-10-08
Last updated2026-09-13 11:57 UTC
CompanyBeckman Coulter
Distributionn/a

Product

DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer

Reason

On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

Identifiers

code_infoREF/UDI-DI: C63522/14987666545089, C63521/14987666545072. Software Versions: SW 1.3.3 and 1.3.4

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0032-2026%22