On 2024-10-16 the FDA classified a Class II recall of BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution provi… by Becton Dickinson And, citing synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoin….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0026-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-09-12 |
| Published | 2024-10-16 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Becton Dickinson And |
| Distribution | US |
Product
BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution providing data management and workflow management functionality across locations where clinical diagnostic activities take place. Catalog Number: 444150
Reason
Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy
Identifiers
| code_info | UDI-DI: 00382904441500 Software versions: 4.20, 4.30, 5.10, 5.20 and 5.30 |
|---|