FDA Device recall Z-0026-2016
Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Esse…
- FDA Device
- Class II
- terminated
- Published 2015-10-14
On 2015-10-14 the FDA classified a Class II recall of Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Esse… by Ortho Solutions, citing ortho Solutions Inc is initiating a recall on behalf of UltOS 3.5mm locking screw. The locking thread feature of the screw was not manufactured to specification. As a consequence….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0026-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-07-14 |
| Published | 2015-10-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ortho Solutions |
| Distribution | CA, CO, GA, IL, LA, NC, NY, TX, WA |
Product
Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Essex CM96FF Rx Only a) Sterile b) Non-Sterile Orthopaedic Bone Screw
Reason
Ortho Solutions Inc is initiating a recall on behalf of UltOS 3.5mm locking screw. The locking thread feature of the screw was not manufactured to specification. As a consequence there is a risk that a screw may not lock onto the plate at the time of primary insertion into the plate and the bone. In the event of this occurring the non functioning screw would need to be replaced with an alternativ
Identifiers
| code_info | Part Number (Lot Number) infomration is as follows: OS421712-NS (lot 7891), OS421714-NS (lot 7891) OS421716-NS (lot 7…Part Number (Lot Number) infomration is as follows: OS421712-NS (lot 7891), OS421714-NS (lot 7891) OS421716-NS (lot 7891, 7520), OS421718-NS (lot 7891, 7520) OS421720-NS (lot 7891), OS421722-NS (lot 7891, 7520) OS421724-NS (lot 7891, 7520), OS421726-NS (lot 7891) OS421728-NS (lot 7891), OS421730-NS (lot 7891) OS421735-NS (lot 7520, 7891) ,OS421740-NS (lot 7520, 7891) OS421745-NS (lot 7520, 7891) ,OS421750-NS (lot 7520, 7891) OS421716 (lot 7519) ,OS421718 (lot 7519) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0026-2016%22