RecallRadar

FDA Device recall Z-0022-2013

3M Steri-Vac Gas Sterilizer/Aerator models 5XL and 8XL. 3M Steri-Vac Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- a…

On 2012-10-17 the FDA classified a Class II recall of 3M Steri-Vac Gas Sterilizer/Aerator models 5XL and 8XL. 3M Steri-Vac Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- a… by 3m Health Care Business, citing 3M determined that the electrical grounding mechanism for the Steri-Vac 5XL and 8XL gas Sterilizer/Aerator did not meet all of the requirements of the standard in that the electri….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0022-2013
ClassificationClass II
Statusterminated
Initiated2012-09-13
Published2012-10-17
Last updated2026-09-13 11:53 UTC
Company3m Health Care Business
DistributionUS

Product

3M Steri-Vac Gas Sterilizer/Aerator models 5XL and 8XL. 3M Steri-Vac Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- and/or moisture-sensitive devices. This gas sterilizer/aerator is intended for indoor use only.

Reason

3M determined that the electrical grounding mechanism for the Steri-Vac 5XL and 8XL gas Sterilizer/Aerator did not meet all of the requirements of the standard in that the electrical wiring, which is insulated and contained within the sealed door should have a stronger connection to the protective electrical grounding mechansim.

Identifiers

code_info
Model 5XL 1-door, serial numbers from 721799 - 722166. Model 5XL 2-door, serial numbers from 820248 - 820285. Mod…Model 5XL 1-door, serial numbers from 721799 - 722166. Model 5XL 2-door, serial numbers from 820248 - 820285. Model 8XL 1-door, serial numbers from 350608 - 350887. Model 8XL 2-door, serial numbers from 450073 - 450105

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0022-2013%22