FDA Device recall Z-0018-2018

Philips Healthcare Ingenuity Core128 X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sect…

On 2017-10-25 the FDA classified a Class II recall of Philips Healthcare Ingenuity Core128 X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sect… by Philips Electronics North America, citing during a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0018-2018
ClassificationClass II
Statusterminated
Initiated2017-01-18
Published2017-10-25
Last updated2026-09-13 11:54 UTC
CompanyPhilips Electronics North America
DistributionUS

Product

Philips Healthcare Ingenuity Core128 X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason

During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.

Identifiers

code_infoSoftware version 4.16