Product Name: OSOM C. Difficile Toxin A/B Test Catalog Numbers: 173 (for United States) and 173 E (for Europe)

On 2013-10-23 the FDA classified a Class II recall of Product Name: OSOM C. Difficile Toxin A/B Test Catalog Numbers: 173 (for United States) and 173 E (for Europe) by Sekisui Diagnostics, citing sekisuki Diagnostics has issued a Product Correction for certain lot numbers of the OSOM C. difficile Toxin A/B Test Kit, because it has been determined that there is potential fo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0015-2014
ClassificationClass II
Statusterminated
Initiated2013-09-16
Published2013-10-23
Last updated2026-09-13 11:53 UTC
CompanySekisui Diagnostics
DistributionUS

Product

Product Name: OSOM C. Difficile Toxin A/B Test Catalog Numbers: 173 (for United States) and 173 E (for Europe). The Test is a qualitative assay (in vitro diagnostic) that employs immunochromatographic, dipstick technology. If C. difficile toxin A or toxin B is present, a blue/gray line will appear in the test line region indicating a positive result. A red control line must appear for the results to be valid. If C. difficile toxins are not present, only the red control line will appear. An invalid test occurs when no control line appears The Test is intended for the qualitative detection of Clostridium difficile toxins A and/or B in human stool samples. This test is intended as an aid in the diagnosis of C.difficile associated disease (CDAD) in patients with symptoms of CDAD.

Reason

Sekisuki Diagnostics has issued a Product Correction for certain lot numbers of the OSOM C. difficile Toxin A/B Test Kit, because it has been determined that there is potential for false negative results.

Identifiers

code_info
The following are part numbers and lot number combinations. PN:173E, LN:121272. PN:173, LN:121273. PN:173, LN:1…The following are part numbers and lot number combinations. PN:173E, LN:121272. PN:173, LN:121273. PN:173, LN:121273A. PN:173E, LN:131155. PN:173, LN:131156.