FDA Device recall Z-0013-2013

Baxter Interlink System Buretrol Solution Set, 117" (3.0 m), with 150 mL Burette (Ball-Valve Drip Chamber) (Non-DEHP)

On 2012-10-17 the FDA classified a Class I recall of Baxter Interlink System Buretrol Solution Set, 117" (3.0 m), with 150 mL Burette (Ball-Valve Drip Chamber) (Non-DEHP) by Baxter Healthcare, citing baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the comple….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0013-2013
ClassificationClass I
Statusterminated
Initiated2012-09-07
Published2012-10-17
Last updated2026-09-13 11:53 UTC
CompanyBaxter Healthcare
DistributionUS

Product

Baxter Interlink System Buretrol Solution Set, 117" (3.0 m), with 150 mL Burette (Ball-Valve Drip Chamber) (Non-DEHP); a sterile Rx IV fluid pathway device; Product Usage: The device is used for the administration of fluids from a container into the patient's vascular system through a vascular access device.

Reason

Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.

Identifiers

code_infoproduct code 2H7519, all lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.