FDA Device recall Z-0012-2026
Allura Xper FD20/20
- FDA Device
- Class II
- ongoing
- Published 2025-10-08
On 2025-10-08 the FDA classified a Class II recall of Allura Xper FD20/20 by Philips Medical Systems Nederland B V, citing the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the sy….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0012-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-09-03 |
| Published | 2025-10-08 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Philips Medical Systems Nederland B V |
| Distribution | US |
Product
Allura Xper FD20/20; Model Numbers: 722038; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Reason
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Identifiers
| code_info | Model Numbers: 722038; UDI-DI: 00884838054226; Serial Numbers: 39, 63, 42, 55, 67, 99, 60, 71, 141, 115, 7, 78, 108, 10…Model Numbers: 722038; UDI-DI: 00884838054226; Serial Numbers: 39, 63, 42, 55, 67, 99, 60, 71, 141, 115, 7, 78, 108, 106, 167, 118, 130, 49, 119, 126, 173, 38, 37, 48, 62, 35, 137, 120, 157, 41, 131, 96, 70, 46, 102, 169, 11, 89, 117, 40, 105, 113, 110, 5, 134, 88, 64, 112, 143, 94, 109, 73, 133, 166, 147, 31, 3, 127, 92, 135, 32, 69, 121, 123, 98, 129, 79, 14, 148, 153, 164, 111, 160, 171, 172, 159, 155, 144, 163, 152, 168, 161, 122, 158, 156, 138, 162, 132, 95, 116, 45, 9, 150, 146, 90, 27, 91, 154, 104, 124, 30, 136, 72, 93, 66, 140, 75, 50, 68, 58, 54, 47, 36, 28, 29; |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0012-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.