FDA Device recall Z-0011-2026
Allura Xper FD20/15 OR Table
- FDA Device
- Class II
- ongoing
- Published 2025-10-08
On 2025-10-08 the FDA classified a Class II recall of Allura Xper FD20/15 OR Table by Philips Medical Systems Nederland B V, citing the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the sy….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0011-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-09-03 |
| Published | 2025-10-08 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Philips Medical Systems Nederland B V |
| Distribution | US |
Product
Allura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Reason
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Identifiers
| code_info | Model Numbers: 722059; UDI-DI: 00884838059122; Serial Numbers: 7, 33, 35, 40, 38, 41, 37, 2; |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0011-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.