FDA Device recall Z-0011-2025
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24
- FDA Device
- Class II
- ongoing
- Published 2024-10-30
On 2024-10-30 the FDA classified a Class II recall of CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24 by Smiths Medical Asd, citing certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0011-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-06-26 |
| Published | 2024-10-30 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Smiths Medical Asd |
| Distribution | n/a |
Product
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24
Reason
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Identifiers
| code_info | UDI/DI 10610586027239, Lot Numbers: 3971709, 3983317, 3983316, 3983315, 3983318, 3983327, 3983320, 3983321, 3983323, 3…UDI/DI 10610586027239, Lot Numbers: 3971709, 3983317, 3983316, 3983315, 3983318, 3983327, 3983320, 3983321, 3983323, 3983328, 3983322, 4013357, 3983325, 3983324, 3983329, 3983326, 3983330, 3983332, 3983331, 3983334, 4013356, 4034028, 4013355, 4046848, 4013353, 4013358, 3983335, 4013360, 4013361, 4013359, 4013363, 4013362, 4013377, 4013374, 4013376, 4013378, 4013373, 4013375, 4034023, 4019430, 4019427, 4019428, 4034024, 4019429, 4034029, 4037751, 4034026, 4037754, 4037753, 4034031, 4034030, 4034025, 4037757, 4046846, 4037752, 4037756, 4040249, 4037755, 4042843, 4040250, 4040251, 4046845, 4042844, 4046847, 4052411, 4048893, 4066532, 4048892, 4048894, 4052410, 4053920, 4052412, 4052413, 4060911 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0011-2025%22