RecallRadar

FDA Device recall Z-0010-2027

Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 24" X 4" 1-LINE (CS 10) YELLOW Model/Catalog Number: 564008 Software Vers…

On 2026-10-07 the FDA classified a Class II recall of Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 24" X 4" 1-LINE (CS 10) YELLOW Model/Catalog Number: 564008 Software Vers… by Asp Global, citing due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0010-2027
ClassificationClass II
Statusongoing
Initiated2026-08-24
Published2026-10-07
Last updated2026-10-13 05:30 UTC
CompanyAsp Global
DistributionUS

Product

Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 24" X 4" 1-LINE (CS 10) YELLOW Model/Catalog Number: 564008 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No

Reason

Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.

Identifiers

code_info
Lot Code: Model Number: 564008 UDI-DI: (01) 00885914044582 All Lots produced from 01SEP2016 to 01MAR2023: S0823STY,…Lot Code: Model Number: 564008 UDI-DI: (01) 00885914044582 All Lots produced from 01SEP2016 to 01MAR2023: S0823STY, S0922STY, S1101STY, S1214STY, R0310STY, R0601STY, R0801STY, R1101STY, Q0301STY, Q0930STY, Q1115STY, P0325STY, P0620STY, P1015STY, N0330STY, N0730STY, M0730STY, M0930STY

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-13 05:30 UTC. Current status: ongoing.

Other recalls by Asp Global

All Asp Global recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0010-2027%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.