eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach

On 2014-10-08 the FDA classified a Class II recall of eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach by Gendex, citing two (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0009-2015
ClassificationClass II
Statusterminated
Initiated2014-08-12
Published2014-10-08
Last updated2026-09-13 11:53 UTC
CompanyGendex
DistributionUS

Product

eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inch reach; eXpert DC, WW, 65 Inch reach; eXpert DC, WW, 55 Inch reach; OEM Arm Assembly w/o cables; Assy, Articulated Arm & Tubehead, eDC; and Assy, Articulated Arm & Tubehead, 765DC Product Usage: intraoral x-ray

Reason

Two (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycle of the arm, prior to the 14lb tube head (full load) being installed onto the assembled articulated arm.

Identifiers

code_info
Models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inc…Models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inch reach; eXpert DC, WW, 65 Inch reach; eXpert DC, WW, 55 Inch reach; OEM Arm Assembly w/o cables; Assy, Articulated Arm & Tubehead, eDC; and Assy, Articulated Arm & Tubehead, 765DC; with multiple part numbers and serial numbers: see distribution list for these details.