stryker EVOLVE, Proline Stem, CoCr, Size: +4, OD: 9.5mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System is a set c…

On 2022-10-12 the FDA classified a Class II recall of stryker EVOLVE, Proline Stem, CoCr, Size: +4, OD: 9.5mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System is a set c… by Wright Medical Technology, citing the outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Prol….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0006-2023
ClassificationClass II
Statusongoing
Initiated2022-08-03
Published2022-10-12
Last updated2026-09-13 11:56 UTC
CompanyWright Medical Technology
DistributionUS

Product

stryker EVOLVE, Proline Stem, CoCr, Size: +4, OD: 9.5mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System is a set configuration designed to address radial head fractures. It combines parts of two systems: EVOLVE Proline and EVOLVE TRIAD.

Reason

The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.

Identifiers

code_infoUDI/DI: (01)008404201856775(11)210831(17)290829 Lot Number 1704255