RecallRadar

FDA Device recall Z-0004-2015

Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applicatio…

On 2014-10-08 the FDA classified a Class II recall of Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applicatio… by Siemens Medical Solutions Usa, citing two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific s….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0004-2015
ClassificationClass II
Statusterminated
Initiated2014-09-02
Published2014-10-08
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: "Querying and retrieving patient history information and/or previous diagnosis and images from other modalities; "X-ray examinations of the urogenital area "Ultrasound examinations "Endourological interventions "Percutaneous interventions "Laparoscopy "Application of fistula "Simple procedures "Extracorporeal shock wave lithotripsy "Uroflow/urodynamics "Pediatric radiological and therapeutic applications

Reason

Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images

Identifiers

code_infoSiemens Uroskop Omnia Max system, material #s: 10762473, serial numbers: 4014 4015

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0004-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.