FDA Device recall Z-0003-2015
Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographi…
- FDA Device
- Class II
- terminated
- Published 2014-10-08
On 2014-10-08 the FDA classified a Class II recall of Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographi… by Siemens Medical Solutions Usa, citing two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific s….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0003-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-09-02 |
| Published | 2014-10-08 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.
Reason
Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
Identifiers
| code_info | Siemens AXIOM Luminos dRF Max system material #s: 10762471 with serial number: 5037. |
|---|