FDA Device recall Z-0003-2014
Cordis ADROIT 6F Guiding Catheter Product Usage: For intravascular introduction of interventional and/or diagnostic devices into the corona…
- FDA Device
- Class II
- terminated
- Published 2013-10-09
On 2013-10-09 the FDA classified a Class II recall of Cordis ADROIT 6F Guiding Catheter Product Usage: For intravascular introduction of interventional and/or diagnostic devices into the corona… by Cordis, citing due to a labeling error related to the conversion of the inner diameter of the catheter from inches to millimeters. The English dimension is correct and is the primary dimension.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0003-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-08-12 |
| Published | 2013-10-09 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Cordis |
| Distribution | US |
Product
Cordis ADROIT 6F Guiding Catheter Product Usage: For intravascular introduction of interventional and/or diagnostic devices into the coronary or peripheral vascular system.
Reason
Due to a labeling error related to the conversion of the inner diameter of the catheter from inches to millimeters. The English dimension is correct and is the primary dimension.
Identifiers
| code_info | Catalog Lots ---------------------------------- 67200000 15918614 67200200 15892736 67200400…Catalog Lots ---------------------------------- 67200000 15918614 67200200 15892736 67200400 15892737, 15897416 67200500 15901314 67203400 15922021 67203600 15897418 67204000 15925687 67205200 15892738, 15892739 67205400 15885645, 15887472, 15897420 67205600 15901319 67206000 15897422 67207200 15892740 67208000 15892741 67208200 15887473, 15887474, 15897424 67208300 15901321 67212600 15916942 67213000 15916943 67227000 15901323 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0003-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.