FDA Device recall Z-0003-2013
IMMAGE Immunochemistry Systems Rheumatoid Factor (RF) Reagent, PN 447070 The IMMAGE Immunochemistry Systems Rheumatoid Factor (RF) reagent,…
- FDA Device
- Class II
- terminated
- Published 2012-10-10
On 2012-10-10 the FDA classified a Class II recall of IMMAGE Immunochemistry Systems Rheumatoid Factor (RF) Reagent, PN 447070 The IMMAGE Immunochemistry Systems Rheumatoid Factor (RF) reagent,… by Beckman Coulter, citing the recall was initiated because Beckman Coulter has confirmed customer reports of QC shifts, lot-to-lot variation, and an increase in the percent of patient samples recovering >2….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0003-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-01-27 |
| Published | 2012-10-10 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Beckman Coulter |
| Distribution | US |
Product
IMMAGE Immunochemistry Systems Rheumatoid Factor (RF) Reagent, PN 447070 The IMMAGE Immunochemistry Systems Rheumatoid Factor (RF) reagent, in conjunction with Beckman Cal 5 Plus, is intended for use in the quantitative determination of human rheumatoid factor concentrations in human serum and plasma samples by rate nephelometry. This assay is designed for use with the IMMAGE Immunochemistry System.
Reason
The recall was initiated because Beckman Coulter has confirmed customer reports of QC shifts, lot-to-lot variation, and an increase in the percent of patient samples recovering >20 IU/mL using the Lots# M012376 and M101865 of IMMAGE RF reagent. The lots identified for the RF reagent might produce false abnormal results.
Identifiers
| code_info | Lot# M012376 and M101865 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0003-2013%22