RecallRadar

FDA Device recall Z-0002-2027

HemoslL AcuStar HIT-lgG(PF-4)

On 2026-10-21 the FDA classified a Class I recall of HemoslL AcuStar HIT-lgG(PF-4) by Instrumentation Laboratory, citing potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0002-2027
ClassificationClass I
Statusongoing
Initiated2026-09-03
Published2026-10-21
Last updated2026-10-27 05:30 UTC
CompanyInstrumentation Laboratory
DistributionUS

Product

HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)

Reason

Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

Identifiers

code_infoPart Number: 0009802028. UDI-DI: 08426950545774. Lot Number (Expiration Date): B40006 (01/16/2027), B38063 (11/01/2026).

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-27 05:30 UTC. Current status: ongoing.

Other recalls by Instrumentation Laboratory

All Instrumentation Laboratory recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0002-2027%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.