FDA Device recall Z-0001-2017
CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor du…
- FDA Device
- Class III
- terminated
- Published 2016-10-12
On 2016-10-12 the FDA classified a Class III recall of CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor du… by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0001-2017 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2015-12-11 |
| Published | 2016-10-12 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
Reason
Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.
Identifiers
| code_info | 6912 , 6913 6914, 6932, 6933 6934, 6922, 6923 6924, CV-6912, CV-6913, CV-6914 Cardiovascular Procedure Kits- Various |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0001-2017%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.