RecallRadar

CPSC recall 02231

Fluke Digital Multimeters

On 2025-10-08 the CPSC announced a recall of Fluke Digital Multimeters by Fluke, citing the recalled units can take longer than normal, up to 18 seconds, to display readings of AC voltages above 500 volts. Users can misinterpret the delayed reading to mean that high….

Read the official CPSC notice

AgencyCPSC
Recall number02231
ClassificationNot classified
Statusunknown
Initiated2002-08-19
Published2025-10-08
Last updated2026-09-13 11:58 UTC
CompanyFluke
DistributionUS

Product

Fluke Digital Multimeters | Multimeters are used to measure voltage, resistance and current. The recalled Fluke digital multimeters measure up to 1,000 volts alternating current (VAC) and direct current (VDC). "Fluke" and "175", "177" or "179" are written on the front of the unit. The tester body is yellow and black, and measures about 7.25 inches long by 3.5 inches wide. Recalled units have a serial number below 79000000. The serial number is written on the back of the instrument, under the hinged stand.

Reason

The recalled units can take longer than normal, up to 18 seconds, to display readings of AC voltages above 500 volts. Users can misinterpret the delayed reading to mean that high voltage is not present. If high voltage is present, users could be exposed to a risk of shock, electrocution, and thermal burns.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: unknown.

Source

Official record: https://www.cpsc.gov/Recalls/2002/CPSC-Fluke-Corp-Announce-Recall-of-Digital-Multimeters

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.