CPSC recall 02218
Greenlee CM-700 and CM-750 electrical meters
- CPSC
- unknown
- Published 2025-10-03
On 2025-10-03 the CPSC announced a recall of Greenlee CM-700 and CM-750 electrical meters by Greenlee Textron Greenlee, citing an incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to provide inaccurate voltage and current readings. A meter that inaccur….
| Agency | CPSC |
|---|---|
| Recall number | 02218 |
| Classification | Not classified |
| Status | unknown |
| Initiated | 2002-08-01 |
| Published | 2025-10-03 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Greenlee Textron Greenlee |
| Distribution | US |
Product
Greenlee CM-700 and CM-750 electrical meters | This recall involves certain CM-700 and CM-750 electrical meters. The 8-inch meters are dark green with bright yellow features. The model number and the words "Greenlee Test Instruments" are printed across the front of the meter. A silver plate on the back of the meter displays the serial number (S/N). Serial numbers included in the recall range from 0203540001 through 0203540650. The electrical meters were made in Taiwan. CM-700 and CM-750 meters outside the above serial number range are not included in this recall.
Reason
An incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to provide inaccurate voltage and current readings. A meter that inaccurately indicates zero voltage or current creates the potential for electric shock or an electrocution hazard from the equipment being tested.
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: unknown.
Source
Official record: https://www.cpsc.gov/Recalls/2002/CPSC-Greenlee-Textron-Inc-Announce-Recall-of-Electrical-Testing-Meters-