Micro X recalls
Micro X appears in 10 official U.S. recall notices indexed by RecallRadar, most recently on 2026-04-29.
Company names are normalized from agency data (legal suffixes removed), so related firms may appear under separate pages.
Recall summary
| Recall notices | 10 |
|---|---|
| Published | 2022-10-05 to 2026-04-29 |
| By agency | FDA Device (10) |
| By classification | Class II (10) |
| By status | ongoing (10) |
| Distribution named in notices | CA, MD, NJ, PR, TX, US, WA |
Recalls
- Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71
FDA Device · Class II · 2026-04-29 · Micro X
Reason: X-Ray generator may malfunction resulting in the X-Ray being inoperable. - Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71, MXU-RV35 Product Description: Rover Mobile X-rays Syste…
FDA Device · Class II · 2025-04-30 · Micro X
Reason: Potential for internal fastening mechanism within generator of mobile x-Ray to fail. - Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number:MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover")…
FDA Device · Class II · 2025-04-30 · Micro X
Reason: Potential for internal fastening mechanism within generator of mobile x-Ray to fail. - Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover"…
FDA Device · Class II · 2025-04-23 · Micro X
Reason: Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode. - Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile X-rays System ("Rover"…
FDA Device · Class II · 2025-04-23 · Micro X
Reason: Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode. - MICRO-X Rover Mobile X-ray System, # MXU-RV19
FDA Device · Class II · 2024-12-18 · Micro X
Reason: During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a v… - MICRO-X Rover Mobile X-ray System, # MXU-RV35
FDA Device · Class II · 2024-12-18 · Micro X
Reason: Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator. - MICRO-X Rover Mobile X-ray System, # MXU-RV35
FDA Device · Class II · 2024-12-18 · Micro X
Reason: Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started. - MICRO-X Rover Mobile X-ray System, # MXU-RV35
FDA Device · Class II · 2024-12-04 · Micro X
Reason: Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106… - MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19
FDA Device · Class II · 2022-10-05 · Micro X
Reason: Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions