- 11278A2 11278AK2 Flexible Ureteroscope Z20615US-BA (08-2018)
FDA Device · Class II · 2022-05-18 · Karl Storz Endoscopy
Reason: Failure to achieve the expected six-log reduction in microorganisms following the disinfection process. - 11272V CMOS VIDEO-CYSTO-URETHROSCOPE REV 11272VA 11272VAK CMOS VIDEO-CYSTO-URETHROSCOPE REV
FDA Device · Class II · 2022-05-18 · Karl Storz Endoscopy
Reason: Failure to achieve the expected six-log reduction in microorganisms following the disinfection process. - Karl Storz Udel Sterilization Tray Instructions for Use, Tray Part Number/Description: 39301ACTS/CAMERA TRAY (STERRAD, INLINE CAMERA HEADS)
FDA Device · Class II · 2022-01-05 · Karl Storz Endoscopy
Reason: To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and maximu… - FIVE S 3.5x65, Flexible Intubation Videoscope, REF: 091361-01 (1 Unit), and 091361-06 (6 Units)
FDA Device · Class II · 2021-11-10 · Karl Storz Endoscopy
Reason: During a regular requalification review, affected flexible intubation video endoscopes did not achieve the required sterility assurance level. There is a risk that the patient may be exposed to a higher risk of infectio… - Flexible Intubation Fiberscope, Part: 11302BD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination…
FDA Device · Class II · 2020-09-02 · Karl Storz Endoscopy
Reason: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with tot… - Flexible Intubation Video Endoscope, Part: 11303BNX, with Instruction Manual: Z20827US - Product Usage: are intended to be used for examina…
FDA Device · Class II · 2020-09-02 · Karl Storz Endoscopy
Reason: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with tot… - Flexible DCI Intubation Fiberscope, Part: 11302BDD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examina…
FDA Device · Class II · 2020-09-02 · Karl Storz Endoscopy
Reason: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with tot… - Flexible Intubation Video Endoscope, Part: 11302BDX, with Instruction Manual: Z18443US - Product Usage: are intended to be used for examina…
FDA Device · Class II · 2020-09-02 · Karl Storz Endoscopy
Reason: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with tot… - Flexible Intubation Fiberscope, Part: 11301BN1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination…
FDA Device · Class II · 2020-09-02 · Karl Storz Endoscopy
Reason: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with tot… - Flexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US - Product Usage: are intended to be used for examina…
FDA Device · Class II · 2020-09-02 · Karl Storz Endoscopy
Reason: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with tot… - Flexible DCI Intubation Fiberscope, Part: 11301BND1, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examina…
FDA Device · Class II · 2020-09-02 · Karl Storz Endoscopy
Reason: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with tot… - Flexible Intubation Fiberscope, Part: 11301AA1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination…
FDA Device · Class II · 2020-09-02 · Karl Storz Endoscopy
Reason: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with tot… - Insufflation tubing to be used with varies insufflator units
FDA Device · Class III · 2020-05-20 · Karl Storz Endoscopy
Reason: The sterile single-use tubing may contain higher ethylene oxide (EO) residues than the acceptable level. - Karl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx series, which includes: 8403AX, AXC, BX,…
FDA Device · Class II · 2020-04-08 · Karl Storz Endoscopy
Reason: The firm has become aware that there is a potential that the glass cover at the distal end of the camera/led lighting on the video laryngoscope could become loose during reprocessing or use. If this occurs during usage,… - Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope
FDA Device · Class II · 2020-04-01 · Karl Storz Endoscopy
Reason: During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the… - Karl Storz, Model # 11161C2, Flexible Neuroscope
FDA Device · Class II · 2020-04-01 · Karl Storz Endoscopy
Reason: During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the… - Karl Storz RHINO-LARYNGO-BRONCHO FIBERSCOPE, Model # 11005BC1
FDA Device · Class II · 2020-04-01 · Karl Storz Endoscopy
Reason: During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the… - Flexible Intubation Fiberscope, Model # 11301AB1
FDA Device · Class II · 2019-09-11 · Karl Storz Endoscopy
Reason: The action is being initiated due to intubation scopes improperly manufactured with a double port luer instead of a single port luer. - FLEXIBLE NEURO-FIBERSCOPE, 3.7MM X 34CM, Model # 11282BN1
FDA Device · Class II · 2019-09-04 · Karl Storz Endoscopy
Reason: The action is being initiated due to Flexible Neuroscopes manufactured with the incorrect channel port. - STORZ High Flow Insufflation Tubing (with filter), Catalog #20400161S/10 (outer box label), ENDOFLATOR 26-430520-1 also referenced on outer…
FDA Device · Class II · 2017-11-22 · Karl Storz Endoscopy
Reason: The label outside the cardboard box is labeled correctly but the pouches inside of some of the boxes were mislabeled as a different product. - Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116…
FDA Device · Class II · 2016-07-13 · Karl Storz Endoscopy
Reason: Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG). - Essential NEO, Audio/Visual Integration Unit An audio/visual integration unit
FDA Device · Class II · 2014-07-16 · Karl Storz Endoscopy
Reason: KARL STORZ Endoscopy-America is recalling the Essential NEO because the connected power cable could potentially become overheated and result in smoke being generated and the cessation of function of the peripheral that…